@ARTICLE{Marzejon_Marcin_Fibre-optic_2019, author={Marzejon, Marcin and Karpienko, Katarzyna and Mazikowski, Adam and Jędrzejewska-Szczerska, Małgorzata}, volume={vol. 26}, number={No 3}, journal={Metrology and Measurement Systems}, pages={561-568}, howpublished={online}, year={2019}, publisher={Polish Academy of Sciences Committee on Metrology and Scientific Instrumentation}, abstract={In this paper, we present a fibre-optic sensor for simultaneous measurement of refractive index and thickness of liquid layers.We designed an experimental low-coherence setup with two broadband light sources and an extrinsic fibre-optic Fabry–Pérot interferometer acting as the sensing head.We examined how the refractive index of a liquid film and its thickness affect spectrum at the output of a fibre-optic interferometer. We performed a series of experiments using two light sources and only one sensing head. The spectra were collected in ranges of 1220÷1340 nm and 1500÷1640 nm. The obtained results show that using two spectra recorded simultaneously for two wavelength ranges enables to determine thickness in a range of 50÷500 μm, and refractive index of a liquid film in a range of 1:00÷1:41 RIU using only one sensing head.}, type={Article}, title={Fibre-optic sensor for simultaneous measurement of thickness and refractive index of liquid layers}, URL={http://czasopisma.pan.pl/Content/113103/PDF/10_MMS_3_INTERNET.pdf}, doi={10.24425/mms.2019.129584}, keywords={fibre-optic, sensors, interferometry, Fabry–Pérot interferometer}, }