Wyniki wyszukiwania

Filtruj wyniki

  • Czasopisma
  • Data

Wyniki wyszukiwania

Wyników: 1
Wyników na stronie: 25 50 75
Sortuj wg:

Abstrakt

The Helmholtz coil constant (k h) is a crucial standard in magnetic moment measurement devices for permanent magnet materials. To overcome the problem of lowaccuracy of the direct-current (DC) calibration method, this study used a constant sinusoidal current in the Helmholtz coil and measured the induced voltage of the detection coil with known coil turns and coil area. Subsequently, the k h was calculated. The noise signal deduction rate in the induction voltage of the detection coil was greater than 99%, its influence on the induction voltage is less than 0.005%, and the repeatability of the calibration results is 0.003% (1δ). The results reveal that the alternating current (AC) method and orthogonal calculation (OC) can accurately measure the valid values of the voltage signal under the influence of the spatial stray field during the calibration of k h.
Przejdź do artykułu

Autorzy i Afiliacje

Kuankuan Zhang
1 2
Mingxing Cao
1
Jian He
1
Wenjie Gong
1
Yunhua Huang
2

  1. Magnetic Materials Measurement Laboratory, National Institute of Metrology, Beijing 100029, China
  2. University of Science and Technology Beijing, Beijing 100083, China

Ta strona wykorzystuje pliki 'cookies'. Więcej informacji